Section 5
Page 73
UVTOCSetup and System Configuration.fm
Setup Menu
1.
Enter the Pre span period by pressing the UP button while in the Pre zero
period screen.
2.
To change the displayed setting, press ENTER.
3.
Asterisks will flash on either side of the setting. Use the UP or DOWN
buttons to increase or decrease the setting (1 to 60 minutes).
4.
When the desired delay time is displayed, press ENTER to place the
value in memory.
5.2.4.3 Clean Period
Clean period is the number of minutes the analyzer is cleaning in autoclean
mode or autocalibration mode if enabled during an autocalibration.
1.
After the Clean period is set, press the UP button while in the Cal delay
screen. The Clean period screen is displayed.
2.
To change the displayed setting, press ENTER.
3.
Asterisks will flash on either side of the setting. Use the UP or DOWN
buttons to increase or decrease the setting (1 to 60 minutes).
4.
When the desired delay time is displayed, press ENTER to place the
value in memory.
5.2.4.4 Purge Period
The Purge period is the number of minutes programmed to purge the
analytical system of any liquid before going online with the sample from the
stream port. Purging the system will ensure that the measurement reflects the
true value of the stream sample.
The purge function is used following an autocalibration, autoclean, single
sample analysis, auto or manual calibration, and validation before signaling
an online status.
1.
Enter the Purge period screen by pressing the UP button while in the
Clean period screen. The Purge delay screen is displayed.
2.
To change the displayed setting, press ENTER.
3.
Asterisks will flash on either side of the setting. Use the UP or DOWN
buttons to increase or decrease the setting. Acceptable values are
2 to 60 minutes.
P r e s p a n p e r i o d
9 m i n
C l e a n p e r i o d
9 m i n
P u r g e p e r i o d
7 m i n
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...