Section 8
Page 141
UVTOCTroubleshooting and Service.fm
Service Menus
Referring to the menus above, the LO/HI indication is an input status, which
will toggle for 0.5 seconds each time a bubble is detected. This is used verify
correct operation during commissioning of an analyzer. The number is the
threshold bubble counts per minute which will cause the associated flow
event, and is different for every analyzer configuration. Sensitivity to flow
failure can be decreased by increasing the number, and increased by
decreasing the number. Setting the threshold to 0, will disable the bubble
counter event completely. The numeric value can be adjusted by pressing
ENTER
and then using the
UP
and
DOWN
keypads. Pressing
ENTER
again
locks the new number. The threshold can be set from 0 (disable) to 200.
Because of the nature of some sample compositions, it is not possible to
guarantee that the default bubble threshold value is always suitable, or even
that the detection method will not cause false triggering. Refer to
for
default bubble threshold values.
Single Stream Operation— When Online, if the analyzer detects more
bubbles per minute than the threshold value, the analyzer will go Offline, with
the pumps still running and the UV lamps off. While Offline, if the analyzer
detects a bubble count per minute less than the threshold, the analyzer will
enter a Purge Period then return to Online mode. If the bubble count is
exceeded in the last minute of the Purge, the analyzer will return to
Offline mode.
Dual Stream Operation— When Online in a stream A, if the analyzer detects
more bubbles per minute than the threshold value, the analyzer will switch to
the other stream, stream B, and enter a purge. Following the Purge Period,
the analyzer will go Online on stream B. On completion of the Purge Period,
the analyzer will revert to the original stream, stream A, enter a Purge and go
online to stream A only if the bubble count is less than the threshold value. If
the bubble count is exceeded the analyzer with revert back to stream B. The
analyzer will thus act appropriately based on the last minute of the Purge
Period for each stream. Note that the UV lamps will be on when Purging is
being done so that if both streams are causing errors the lamps are on
continually. For this reason, this condition should not be ignored.
Purge Period Operation— All bubble issues are ignored during the Purge
Period except for the last minute of the Purge period. The analyzer will
continue to cause an event based on the bubble count/ bubble threshold
values during the last minute of the Purge Period.
S p a r g e r f l o w
L O
15
U V f l o w
H I
30
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...