Page 96
Calibration Menu
UVTOCOperation.fm
Section 6
6.9.1 Performing a New Validation
1.
If a new validation operation is required, press ENTER. This displays the
TOC standard last used to perform a validation.
2.
The blinking asterisks on either side of the displayed value indicate that it
can be changed. Press ENTER to move to the prompt screen and ensure
the standard is connected to the calibration port.
3.
Press ENTER. The solution taken from the calibration standard port is
analyzed. The value increases as the analysis progresses and a
countdown timer is displayed.
The validation continues until completed, after which it displays the normal
validation screen for ten seconds.
An alternate validation percentage screen is displayed following the mg/L
screen, as shown below.
6.10 Calibration Menu
6.10.1 Calibration Overview and Options
The analyzer supports primary calibration, manual calibration, automatic
calibration, and multi-point calibration.
Primary and Manual calibrations are identical in operation. A primary
calibration should be done after peristaltic tube or lamp replacement.
In this manual, TOC is expressed in
mg/L while CO
2
concentrations are
expressed in ppm.
The analyzer is calibrated by measuring and storing the CO
2
gas ppm
produced by at least two known TOC standards. A calibration set may be
represented by at least two points on a graph of ppm CO
2
versus a zero and a
span (mg/L TOC) so that a straight line interpolation and extrapolation may be
done between the points. The lower point is defined by the gas ppm x
1
produced when DI water is applied to the analyzer. This defines the point
V a l i d a t i o n
* 1 0 0 0 m g / l *
A p p l y S T D t o
c a l i b r a t i o n p o r t
V a l i d a t i o n
9 9 7 m g / l
99s
2 / 1 5 / 2 0 0 0
12 :3 8
V a l
9 9 7 m g / l
2 / 1 5 / 2 0 0 0
12 :3 8
V a l 9 9. 7%
Содержание 1950Plus
Страница 2: ...Page 2 4195M0040title page fm...
Страница 12: ...Page 12 Safety Precautions UVTOCSafety Precautions fm...
Страница 24: ...Page 24 EPA Mode UVTOCIntroduction fm...
Страница 26: ...Page 26 Installation UVTOCInstallation Stopper fm...
Страница 28: ...Page 28 Unpacking and Installation UVTOCUnpacking Installation Setup fm Section 2 Figure 6 Analyzer Dimensions...
Страница 54: ...Page 54 Storage and Handling Tips UVTOCReagents and Standards fm...
Страница 60: ...Page 60 Analyzer Adjustments UVTOCCommissioning fm...
Страница 62: ...Page 62 Operation UVTOCOperation Stopper fm...
Страница 110: ...Page 110 Maintenance UVTOCMaintenance Stopper fm...
Страница 128: ...Page 128 Replacing the Fuses UVTOCMaintenance fm...
Страница 137: ...Section 8 Page 137 UVTOCTroubleshooting and Service fm Service Menu Structure 8 5 Service Menu Structure...
Страница 146: ...Page 146 Fault ID Codes UVTOCTroubleshooting and Service fm...
Страница 152: ...Page 152 TC Conversion UVTOC_TC_conversion fm...
Страница 160: ...Page 160 Analyzer Pump Configurations UVTOC_PumpConfiguration fm...
Страница 164: ...Page 164 Glossary UVTOCGlossary fm...
Страница 172: ...Page 172 Replacement Parts UVTOCReplacement Parts fm...
Страница 176: ...Page 176 General Information UVTOCGeneral Information Stopper fm...
Страница 182: ...Page 182 4195M0040Certification fm Certification...