Chapter 3
85
Setting Measurement Conditions
Setting the Scales
3
. Set
ting
M
e
as
ur
em
e
nt
C
onditions
Setting the Scales
Auto scale
The auto scale function is used to tailor each scale (scale/division and the reference line
value) automatically in such a way that traces will appear at the proper size on the screen
for easy observation.
Single Trace Auto Scale
Follow the procedure below to perform the auto scale function on a specific trace.
Step 1.
Press
(or
) and
(or
) to select the trace for
which the auto scale function will be performed.
Step 2.
Press
.
Step 3.
Press
Auto Scale
.
Auto Scale on All Traces Within a Channel
Step 1.
Press
(or
) to select the channel for which the auto scale function
will be performed.
Step 2.
Press
.
Step 3.
Press
Auto Scale All
.
Manual scale adjustment on a rectangular display format
For a rectangular display format, four parameters are used to manually adjust the scales
(Table 3-3 and Figure 3-11).
Table 3-4
Adjustable scale features on a rectangular display format
Adjustable feature
Description
Divisions (
Divisions
)
Defines the number of divisions on the Y-axis. An even number
from 4 to 30 must be used. Once set, it is commonly applied to
all traces displayed in any rectangular format within that
channel.
Scale/Division (
Scale/Div
)
Defines the number of increments per division on the Y-axis.
The value applies only to the active trace.
Reference position
(
Reference Position
)
Defines the position of the reference line. The position must be
specified using the number assigned to each division on the
Y-axis starting at 0 (the least significant) running up to the
number of divisions being used (the most significant). The
position applies only to the active trace.
Reference line value
(
Reference Value
)
Defines the value corresponding to the reference line. It must be
set using the unit on the Y-axis. The reference line value applies
only to the active trace.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...