Chapter 16
535
Specifications and Supplemental Information
Corrected System Performance
16. S
pecifica
tions and
Su
pple
m
en
ta
l Infor
m
ation
Corrected System Performance
The specifications in this section apply for measurements made with the Agilent
E5070B/E5071B Network Analyzer under the following conditions:
•
No averaging applied to data
•
Environmental temperature of 23
°
C
±
5
°
C, with less than 1
°
C deviation from the
calibration temperature
•
Response and isolation calibration not omitted
Table 16-1
System Dynamic Range
Description
Specification
Supplemental Information
System Dynamic Range
*1*2
300 kHz to 3 MHz
IF bandwidth = 3 kHz
85 dB
3 MHz to 1.5 GHz
95 dB
98 dB
1.5 GHz to 3 GHz
97 dB
100 dB
3 GHz to 4 GHz
96 dB
99 dB
4 GHz to 6 GHz
92 dB
94 dB
6 GHz to 7.5 GHz
87 dB
90 dB
7.5 GHz to 8.5 GHz
80 dB
83 dB
300 kHz to 3 MHz
IF bandwidth = 10 Hz
110 dB
3 MHz to 1.5 GHz
120 dB
123 dB
1.5 GHz to 3GHz
122 dB
125 dB
3 GHz to 4 GHz
121 dB
124 dB
4 GHz to 6 GHz
117 dB
119 dB
6 GHz to 7.5 GHz
112 dB
115 dB
7.5 GHz to 8.5 GHz
105 dB
108 dB
*1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source
maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals
into account.
*2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.25GHz due to spurious receiver
residuals.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...