![Agilent Technologies E5070B Скачать руководство пользователя страница 83](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065083.webp)
Chapter 3
77
Setting Measurement Conditions
Setting Stimulus Conditions
3
. Set
ting
M
e
as
ur
em
e
nt
C
onditions
Setting power range manually
By turning off the Auto Power Range set function, you can set the power level at the
frequency sweep independently for each test port manually, within the range of
−
20 dBm
to 10 dBm (for instruments with option 214, 314, or 414, the range of
−
55 dBm to 10 dBm)
and at a resolution of 0.05 dB.
Selecting range
When option 214, 314, or 414 is installed, you can select from the following power ranges.
Selectable power ranges
−
20 dBm to 10 dBm
−
25 dBm to 7 dBm
−
30 dBm to 2 dBm
−
35 dBm to
−
3 dBm
−
40 dBm to
−
8 dBm
−
45 dBm to
−
13 dBm
−
50 dBm to
−
18 dBm
−
55 dBm to
−
23 dBm
NOTE
The range selection is common to all ports: You cannot select it for each port
independently.
The procedure for selecting the range is as follows:
Step 1.
Press
or
to select the channel for which you want to set the
power range.
Step 2.
Press
.
Step 3.
Press
Power
.
Step 4.
Turn OFF
Auto Range
.
Step 5.
Press
Power Ranges
.
Step 6.
Press the desired softkey to select the power range.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...