530
Chapter 15
Measurement Examples
Executing Power Calibration
2. Setting Address of Power Meter
Follow these steps to configure the power meter's GPIB address.
Setting Description
Key Operation
GPIB address of the power meter: 14
-
Misc Setup
-
GPIB Setup
-
Power Meter
Address
-
3. Setting Stimulus Condition
Follow these steps to configure the stimulus conditions:
Setting Description
Key Operation
Presetting
-
OK
Sweep type: Power
-
Sweep Type
-
Power Sweep
Fixed frequency: 1 GHz
-
Power
-
CW Freq
-
Power range: -40 dBm to -8 dBm
-
Power
-
Power Ranges
-
-40 to -8
Start value: -40 dBm
Stop value: -10 dBm
Number of points: 61
-
Points
-
4. Executing Zero Adjustment and Calibration of Power Sensor
Execute the zero adjustment and calibration of the power sensor according to the
E4418B
Power Meter User's Guide
.
5. Setting Calibration Data Measurement Conditions
Follow these steps to set the calibration data measurement conditions:
Setting Description
Key Operation
Selecting the test port: 1
-
Power Calibration
-
Select Port
-
1
Selection of power sensor: A
-
Power Calibration
-
Use Sensor [ A ]
Number of measurements at one
measurement point: 4
-
Power Calibration
-
Num of Readings
-
Tolerance during power calibration: 5 dB
- Power Calibration - Tolerance -
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...