Chapter 4
155
Calibration
Calibration between Ports of Different Connector Types
4. Cal
ibra
tion
Calibration between Ports of Different Connector Types
When you perform calibration between ports of different connector types, you need to use
a different calibration kit for each test port. In addition, for transmission measurement
between 2 ports, you need to use adapters suitable for the connector types of both ports.
For example, in order to perform full 2-port calibration between port 1 of an N-type
connector and port 2 of a 3.5-mm connector, you need to use an N-type connector
calibration kit (for example, 85032F) for reflection measurement of port 1, a 3.5-mm
connector calibration kit (for example, 85033E) for reflection measurement of port 2, and
an N-3.5-mm adapter for transmission measurement between ports 1 and 2.
Because you cannot use a different calibration kit for each port in normal calibration of the
E5070B/E5071B, you need to use the following VBA macro to perform calibration
between ports of different connector types.
Storage folder
VBA macro name (project name)
D:\Agilent
AdapterCharacterization.vba
NOTE
Never delete this VBA macro, since even system recovery cannot restore it.
This VBA macro lets you select a calibration kit for each test port and each pair of test
ports when performing calibration. In addition, it lets you select any adapter (2-port
Touchstone file) whose characteristics have been determined for a standard between test
ports.
NOTE
This VBA macro has an adapter characterization function to obtain the characteristics of an
adapter and save them to a 2-port Touchstone file. For more information, see “Adapter
Characterization” on page 160.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...