96
Chapter 4
Calibration
Measurement Errors and their Characteristics
Directivity error (Ed)
Directivity errors are caused by the fact that, in a reflection measurement, signals other
than the reflection signal from the DUT are received by receiver T1 (Figure 4-1) through
the directivity coupler. When a certain port is a stimulus port, this error can be defined as a
constant value for each stimulus port because the state of the termination at the other ports
does not change. The number of directivity errors of the E5070B/E5071B is the number of
stimulus ports you use.
Ed1
Directivity error of port 1
Ed2
Directivity error of port 2
*1.Options 313, 314, 413, and 414 only
Ed3
*1
Directivity error of port 3
*2.Options 413 and 414 only
Ed4
*2
Directivity error of port 4
Isolation error (Ex)
An isolation error (crosstalk error) is caused by signals other than the transmission signal
of the DUT leaking to the test receiver of the transmission measurement port in
transmission measurements. When a certain port is a stimulus port, an isolation error is
defined for each of the other ports. Therefore, the number of isolation errors for the
E5070B/E5071B is the total number of combinations of stimulus ports and response ports.
Ex21, Ex31
*1
, and Ex41
*1.Options 313, 314, 413, and 414 only
*2
*2.Options 413 and 414 only
Isolation error when port 1 is a stimulus port
Ex12, Ex32
Isolation error when port 2 is a stimulus port
, and
Isolation error when port 3 is a stimulus port
, and
Isolation error when port 4 is a stimulus port
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...