![Agilent Technologies E5070B Скачать руководство пользователя страница 330](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065330.webp)
324
Chapter 8
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Figure 8-10
Connection of measured mixer
NOTE
If two or more spectrum components are involved at a similar level, it is necessary to use a
filter in order to prevent unnecessary spectrum components from entering the receiver port;
in this way, you can prevent the total magnitude from reaching a damage level of 6.3 Vp-p.
Procedure for setting measurement parameters
Follow the steps below to set the measurement parameters. In this case, we measure each
parameter of magnitude, phase, and group delay in conversion loss (S12).
Step 1.
Press
(or
) to activate the channel you want to measure.
Step 2.
Press
to set
Num of Traces
to
3
.
Step 3.
Press
Allocate Traces
to select the Graph Layout (
).
Step 4.
Press
to set the measurement parameter to
S12
.
Step 5.
Press
to set
LogMag
(magnitude) to Trace 1.
Step 6.
Press
to select Trace 2.
Step 7.
Press
to set the measurement parameter to
S12
.
Step 8.
Press
to set
Phase
(phase) to Trace 2.
Step 9.
Press
to select Trace 3.
Step 10.
Press
to set the measurement parameter to
S12
.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...