Chapter 4
231
Calibration
Vector-Mixer Calibration
4. Cal
ibra
tion
Set the stimulus conditions for the channel you want to calibrate. For the necessary steps,
refer to “Setting Stimulus Conditions” on page 70.
You must also set the external signal source in advance. For the necessary steps, refer to “2.
Setting External Signal Source” on page 316.
NOTE
If you characterize a calibration mixer with an IF filter, we recommend that you perform
full 4-port calibration in advance, since it simplifies the evaluation procedures. For detailed
information on full 4-port calibration, see “Full 4-Port Calibration” on page 124.
Step 1.
Press
.
Step 2.
Press
Load & Project
.
Step 3.
Specify “D:\Agilent\MixerCharacterization.vba” for the file name in the Open dialog box
and then press the
Open
button.
Step 4.
Press
to run the macro (see Figure 4-72).
Step 5.
Select
Port 1
(1 in Figure 4-72) to characterize the calibration mixer 1 with IF filter. In this
case, the data are saved to a temporary file (MIXER_1.s2p).
NOTE
For detailed information on characterizing the calibration mixer, see “Characterizing
procedure for calibration mixer (with IF filter)” on page 225.
Step 6.
Select
Figure 4-72) to characterize calibration mixer 2 with IF filter. Here, the
data are also saved to a temporary file (MIXER_2.s2p).
NOTE
For detailed information on characterizing the calibration mixer, see “Characterizing
procedure for calibration mixer (with IF filter)” on page 225.
Step 7.
Press
(Optional) Balanced Mixer
(2 in Figure 4-72).
Figure 4-72
Vector Mixer Characterization Macro
Step 8.
As the Vector Mixer Characterization Macro is running, the data files of the pre-measured
calibration mixer with IF filter (MIXER_1.s2p, MIXER_2.s2p) are read automatically into
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...