Calibration
User-characterized ECal
6. Storing the User Characteristics to the Memory
Step 1.
When the measurement is complete, the User Characteristic Store screen shown in Figure
4-49 appears.
Specify a user number (a location number in the memory where you want to store the user
characteristics) using 1 in Figure 4-49. If the specified location number is not used for
storage, the parts Characterization, Connectors, and Adapter Description are left blank; if
already used, the stored contents are displayed.
Figure 4-49
User Characteristic Store screen
Step 2.
Enter the following information.
Characterization
Enter the information (operator, used analyzer, and so on) when
measuring user characteristics as necessary.
Connectors
Select the connector types of the adapters for the ECal module’s test
ports. Male and female in the list of connected types indicate male and
female adapter, respectively. Select “No adapter” if no adapter is used on
a port.
Adapter Description
Enter the detailed information on the adapters connected to each port as
necessary.
The information you have entered is displayed when checking the user characteristics
information by using the key strokes:
-
ECal
-
Characterization Info
.
Step 3.
Press
Write
(5 in Figure 4-49).
At this time, if user characteristics are already stored for the specified user number, a
dialog appears to confirm overwriting. Click the
OK
button.
NOTE
Although the maximum number of user characteristics stored to the ECal’s memory is
usually five, this number may be limited by memory size because the size of
user-characteristics data is not fixed and increases in proportion to the number of
measurement points. An error occurs when the
Write
button is pressed if the total size
added the new user characteristics exceeds this limitation due to memory size.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...