114
Chapter 4
Calibration
1-Port Calibration (reflection test)
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
1-Port Cal
.
Step 5.
Press
Select Port
.
Step 6.
Select a test port (and corresponding S parameter) on which 1-port calibration will be
performed.
Softkey
Function
1
Selects port 1
2
Selects port 2
3
Selects port 3
4
Selects port 4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
calibration coefficient (pressing
Done
).
Step 7.
Connect an OPEN calibration standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Step 8.
Press
Open
to start the calibration measurement.
Step 9.
Connect a SHORT calibration standard to the test port (connector to which the DUT is to
be connected) selected in Step 6.
Step 10.
Press
Short
to start the calibration measurement.
Step 11.
Connect a LOAD calibration standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Step 12.
Press
Load
to start the calibration measurement.
Step 13.
Press
Done
to terminate the 1-port calibration process. Upon pressing this key, calibration
coefficients will be calculated and saved. The error correction function will also be
automatically enabled.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...