Chapter 7
295
Fixture Simulator
Evaluating Balanced Devices (balance-unbalance conversion function)
7
. F
ixt
ur
e
Sim
ula
to
r
Measurement parameters of balanced devices
Turn on the balance-unbalance conversion function to measure the following parameters.
R
Mixed mode S-parameter
R
Imbalance parameter
R
CMRR (Common Mode Rejection Ratio)
Mixed mode S-parameter
By turning on the balance-unbalance conversion function, you can obtain the S-parameter
of the balanced port separately for two modes, the differential mode and the common
mode. Figure 7-9 shows the notation of the S-parameter in balance measurement (mixed
mode S-parameter).
Figure 7-9
Notation of mixed mode S-parameter
Figure 7-10 to Figure 7-12 show the mixed mode S-parameter when measuring each
balanced device.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...