![Agilent Technologies E5070B Скачать руководство пользователя страница 324](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065324.webp)
318
Chapter 8
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Step 5.
Select the external signal source you want to use.
NOTE
Switching Time
allows you to set the wait time after setting the external signal source’s
frequency and power in ms.
Procedures for setting external signal source frequency
The frequencies set to the external signal source should be a multiplier (M
LO
), divisor
(D
LO
), and offset (O
LO
) for the basic frequency setting as shown in the following formula
(refer to Figure 8-4 or Figure 8-5):
<LO Frequency> = <Basic Frequency>
×
M
LO
/ D
LO
+ O
LO
Figure 8-4
Example of external signal source frequency setting (swept IF measurement)
Figure 8-5
Example of external signal source frequency setting (fixed IF measurement)
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...