Appendix D
697
Softkey Functions
Marker Function Menu
D. Sof
tke
y
Fu
nctions
Marker Function Menu
Key Operation
Function
SCPI Command
Displays softkeys for setting the sweep range using markers and other marker options.
Marker
→
Start
Sets the starting frequency to the stimulus value of the active marker on the
active trace. Even if the reference marker is enabled and a relative stimulus
value is displayed, the absolute value will be used.
:CALC{1-16}:MARK{1-10}:SET STAR
Marker
→
Stop
Sets the ending frequency to the stimulus value of the active marker on the
active trace. Even if the reference marker is enabled and a relative stimulus
value is displayed, the absolute value will be used.
:CALC{1-16}:MARK{1-10}:SET STOP
Marker
→
Center
Sets the center frequency to the stimulus value of the active marker on the
active trace. Even if the reference marker is enabled and a relative stimulus
value is displayed, the absolute value will be used.
:CALC{1-16}:MARK{1-10}:SET CENT
Marker
→
Reference
Sets the value of the reference line to the response value of the active
marker on the active trace. A softkey having the same function is also
provided in “Scale Menu” on page 713.
:CALC{1-16}:MARK{1-10}:SET RLEV
Marker
→
Delay
Sets the electrical delay to the value of the group delay at the position of the
active marker (a value smoothed with the aperture of 20% regardless of the
setting of smoothing).
:CALC{1-16}:MARK{1-10}:SET DEL
Discrete
Enables or disables the discrete marker function. When enabled, the marker
will be placed at the nearest measurement point from the specified marker
stimulus value. When disabled, the marker will be placed at the point
corresponding to the specified marker stimulus value (by interpolating gaps
between measurement points).
:CALC{1-16}:MARK{1-10}:DISC
Couple
Enables or disables marker coupling. When enabled, marker settings and
movements become effective for all traces on a channel. When disabled,
marker settings and movements are done independently on each trace.
:CALC{1-16}:MARK{1-10}:COUP
Marker Table
Enables or disables the marker table display. The marker table lists the
values of markers for all traces defined on a channel using the bottom third
of the screen. If a channel holds too many markers to be displayed in the
display area, use the scroll bar on the right to view the entire table.
:DISP:TABL
:DISP:TABL:TYPE MARK
Statistics
Enables or disables the display of statistics. When enabled, the following
three statistical data items (Mean value, Standard deviation, Peak-to-peak)
will be displayed on the screen. For details, see “Determining the Mean,
Standard Deviation, and p-p of the Trace” on page 272.
:CALC{1-16}:MST
Annotation Options
Displays softkeys for the appearance of the marker values displayed on the screen.
Marker Info X Pos
Specifies the horizontal display position by the width of the display area as
a percentage.
:DISP{1-16}:WIND:TRAC{1-16}:ANN:
MARK:POS:X
Marker Info Y Pos
Specifies vertical display position by the height of the display area as a
percentage.
:DISP{1-16}:WIND:TRAC{1-16}:ANN:
MARK:POS:Y
Align
Enables or disables the mode to align the display position of the marker
value for each trace to trace 1.
:DISP:WIND{1-16}:ANN:MARK:ALIG
Active Only
Toggles on/off the display of marker values for active traces only.
:DISP:WIND{1-16}:ANN:MARK:SING
Return
Returns to the “E5070B/E5071B Menu (Top Menu)” on page 646.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...