![Agilent Technologies E5070B Скачать руководство пользователя страница 611](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065611.webp)
Appendix A
605
Manual Changes
Manual Changes
A. Manual
Changes
Change 4
The firmware revision A.03.54 or below does not support the following functions. Please
delete the descriptions about these functions in this manual.
o Auto Power Range set function
Change 3
The equipments that firmware revision at the factroy shipment is A.03.53 or below is not
eligible for the following specs. support the following functions. Please refer to original
attached manuals about them.
o Initial Source Port Control function
Change 2
Figure 16-4 dimensions of rear view was changed. Eligible serial numbers are
MY42100211 (for E5070B), MY42100407 (for E5071B) and later. For the earlier serial
number equipment, please refer to original attached manuals.
Change 1
The firmware revision A.03.0x does not support the following functions. Please delete the
descriptions about these functions in this manual
o User Characterized ECal
o Confidence check on calibration coefficients using ECal
o 4-port network embedding/de-embedding (Fixture simulator)
300 kHz to 3 MHz
(source power level = +10 dBm)
0.035
°
rms (23
°
C
±
5
°
C, typical)
0.05
°
rms (high temperature
mode: ON, typical)
3 MHz to 4.25 GHz
(source power level = +10 dBm)
0.007
°
rms (23
°
C
±
5
°
C, typical)
0.02
°
rms (high temperature
mode: ON, typical)
4.25 GHz to 6 GHz
(source power level = +8 dBm)
0.021
°
rms (23
°
C
±
5
°
C, typical)
0.035
°
rms (high temperature
mode: ON, typical)
6 GHz to 7.5 GHz
(source power level = +6 dBm)
0.021
°
rms (23
°
C
±
5
°
C, typical)
0.035
°
rms (high temperature
mode: ON, typical)
7.5 GHz to 8.5 GHz
(source power level = +6 dBm)
0.035
°
rms (23
°
C
±
5
°
C, typical)
0.05
°
rms (high temperature
mode: ON, typical)
*1.Trace noise is defined as a ratio measurement of a through, at IFBW = 3 kHz.
Table A-7
Test Port Input (Trace Noise)
Description
Specification
Supplemental Information
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...