499
Measurement Examples
Evaluating a Duplexer
15
. Me
asur
eme
n
t Exa
m
ple
s
Figure 15-14
Measurement result (segment display: order base)
9. Analyze the Parameters
The parameters for the duplexer are determined.
Step 1.
Determine the insertion loss and 3-dB bandwidth for Tx.
Setup Description
Key Operation
Marker coupling: OFF
-
Couple
(Turn it
OFF
)
Activating Trace 1
Marker 1: ON
Search/Tracking: ON
-
Tracking
(Turn it
ON
)
Moving the marker to the trace maximum
Max
Bandwidth search: ON
Bandwidth
(Turn it
ON
.)
In the example shown in Figure 15-15, insertion loss (
loss
) is 1.243 dB, and the 3-dB
bandwidth (
BW
) is 85.53 MHz.
Step 2.
Determine the insertion loss and 3-dB bandwidth for the Rx.
Setup Description
Key Operation
Activating Trace 2
Marker 1: ON
Search/Tracking: ON
-
Tracking
(Turn it
ON
.)
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...