![Agilent Technologies E5070B Скачать руководство пользователя страница 356](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065356.webp)
350
Chapter 9
Analysis in Time Domain (Option 010)
Characteristics of Response in Time Domain
Identifying mismatch type
The transformation in the low pass mode simulates the response in the TDR measurement.
In addition to mismatch locations, the response includes information on the mismatch type.
Figure 9-12 shows each mismatch type and the response waveform corresponding to it. In
the low pass mode, you can simulate the response of the step signal and the impulse signal.
From the viewpoint of mathematics, the response of the impulse signal is the waveform
obtained by differentiating the response of the step signal.
Figure 9-12
Mismatch type and response in low pass mode
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...