![Agilent Technologies E5070B Скачать руководство пользователя страница 147](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065147.webp)
Chapter 4
141
Calibration
3-port TRL calibration
4. Cal
ibra
tion
Figure 4-33
Connection of standards in 3-port TRL calibration
Operational procedure
NOTE
The shapes of actual standards differ from those shown in Figure 4-33 which are just
symbols for illustration.
Step 1.
Press
or
to select the channel for which you want to execute
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
3-Port TRL Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Select test ports for which you want to execute TRL 3-port calibration. (In the following
procedure, test ports you select here are expressed as test ports x, y, and z.)
Softkey
Function
1-2-3
Selects test ports 1, 2, and 3.
1-2-4
Selects test ports 1, 2, and 4.
1-3-4
Selects test ports 1, 3, and 4.
2-3-4
Selects test ports 2, 3, and 4.
NOTE
An asterisk (*) in the upper right of the softkey indicates that a calibration coefficient exists
that will be cleared if you select the test port and execute acquisition of the calibration
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...