410
Chapter 12
Optimizing Measurements
Improving Measurement Throughput
Improving Measurement Throughput
This section explains the following three methods to improve the measurement throughput.
•
“Using Fast Sweep Modes” on page 410
•
“Turning off the updating of information displayed on the LCD screen” on page 414
•
“Turning off system error correction” on page 414
Using Fast Sweep Modes
The E5070B/E5071B provides four sweep modes: “Stepped mode” and “swept mode,” and
their accelerated versions “fast stepped mode” and “fast swept mode.” You can shorten
sweep time as shown in Table 12-1 by using the swept mode or its fast modes.
NOTE
If it is not necessary to shorten the sweep time, use the stepped mode (preset
configuration).
Table 12-1
Sweep time criteria (each measurement point interval)
Measurement
point interval
Shorter (faster sweep)
→
Longer (slower sweep)
2 MHz or less
Swept
Fast swept
→
Fast stepped
→
Stepped
2 MHz to 5 MHz
Fast swept
→
Swept
→
Fast stepped
→
Stepped
5 MHz to 8 MHz
Fast swept
→
Swept
Fast stepped
→
Stepped
8 MHz to 10 MHz
Fast swept
→
Fast stepped
→
Stepped
Swept
10 MHz to 30 MHz
Fast swept
→
Fast stepped
→
Stepped
→
Swept
30 MHz to 50 MHz
Fast swept
Fast stepped
→
Stepped
→
Swept
50 MHz to 70 MHz
Fast stepped
→
Fast swept
Stepped
→
Swept
70 MHz or more
Fast stepped
→
Stepped
→
Fast swept
→
Swept
Table 12-2
Characteristics of sweep mode
Stepped
Swept
Fast stepped
Fast swept
Measurement
reliability
High
Relatively low
Relatively low
Low
Sweep time
Long
(Except when the
measurement point
interval is
approximately 10
MHz or more)
Short
(Only when the
measurement point
interval is
approximately 10
MHz or less)
Short
(Shortest when the
measurement point
interval is
approximately 30
MHz or more)
Short
(Shortest when the
measurement point
interval is
approximately 30
MHz or less)
Restrictions
on
measurement
No particular
restriction
DUTs with long
electrical delay
cannot be measured
correctly
No particular
restriction
DUTs with long
electrical delay
cannot be measured
correctly
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...