Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Procedures for controlling external signal source
Follow the steps below to control an external signal source.
Step 1.
Press
-
Frequency Offset
to display the “Frequency Offset Menu.”
Step 2.
Press
External Source
to display the “External Source Menu.”
Step 3.
Press
Frequency Format
to turn it
ON
. The frequency setting of the external signal source
(LO) is displayed as shown in Figure 8-6.
Figure 8-6
Display of external signal source setting
Step 4.
Press
Control
to turn the external signal source control
ON
, which allows you to send
setting values such as frequency to the external signal source and start the control. If the
external signal source control is turned
ON
, (Ctrl) will be displayed on the screen as shown
Figure 8-7
Display of external signal source control
NOTE
If the external signal source control fails, the GPIB setting may be wrong. See "Setting the
GPIB" on page 426.
3. Avoid Spurious Function
The E5070B/E5071B has the avoid spurious function (Figure 8-8). This function allows
you to implement measurement by avoiding the spurious signal generated from the
following signals:
•
Stimulus signal and its harmonics
•
Local signal within the receiver and its harmonics
•
For an external signal source (if
LO Frequency
is turned
ON
), the set signal and its
harmonics
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...