590
Appendix A
Manual Changes
Manual Changes
Figure A-5
Performing 2-port TRL/LRM calibration
Step 2.
Press the
Thru
button (1 in Figure A-6).
Step 3.
Make the through condition for test ports 1 and 2 and press the
[ ]Thru 1-2
button (2 in
Figure A-6). The through measurement is performed, and an asterisk appears at the
[ ]Thru
1-2
button and then a check mark [v] also appears on the
Thru
button.
Figure A-6
Example of performing 2-port TRL/LRM calibration
Step 4.
Press the
Reflection
button (3 in Figure A-6).
Step 5.
Connect the Reflection standard to test port 1 and press the
[ ]Reflection1
button. The
reflection measurement is performed, and an asterisk appears at the
[ ]Reflection1
button.
Step 6.
Connect the Reflection standard to test port 2 and press the
[ ]Reflection2
button. The
reflection measurement is performed, and an asterisk appears at the
[ ]Reflection2
button
and then a check mark [v] also appears on the
Reflection
button.
Step 7.
Press the
Match
button (4 in Figure A-6).
Step 8.
Connect the Match standard to test port 1 and press the
[ ]Match1
button. The match
measurement is performed, and an asterisk appears at the
[ ]Match1
button.
Step 9.
Connect the Match standard to test port 2 and press the
[ ]Match2
button. The reflection
measurement is performed, and an asterisk appears at the
[ ]Match2
button and then a
check mark [v] also appears on the
Match
button.
Step 10.
Press the
Line1
button (5 in Figure A-6).
Step 11.
Connect the Line 1 standard between test ports 1 and 2 and press the
[ ]Line1 1-2
button.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...