![Agilent Technologies E5070B Скачать руководство пользователя страница 213](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065213.webp)
Chapter 4
207
Calibration
Specifying Different Standard for Each Frequency
4. Cal
ibra
tion
Example 4-4
When a part of a frequency range specified with a subclass lies outside the
measurement frequency range
For a measurement frequency range (1 GHz - 6 GHz), if you define a standard of subclass
1 (1 GHz - 5 GHz) and a standard of subclass 2 (5 GHz-8 GHz), a part of subclass 2 (6
GHz-8 GHz) lies outside the measurement frequency range, but the other part (5 GHz-6
GHz) lies within it, so calibration is possible.
Figure 4-60
When a part of a frequency range lies outside the range
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...