![Agilent Technologies E5070B Скачать руководство пользователя страница 116](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065116.webp)
110
Chapter 4
Calibration
THRU Response Calibration (transmission test)
THRU Response Calibration (transmission test)
In THRU response calibration, calibration data are measured by connecting a THRU
standard to the desired test port. This calibration effectively eliminates the frequency
response transmission tracking error from the test setup in a transmission test using that
port (Figure 4-7). It is also possible to carry out an isolation calibration using a LOAD
standard in the process of THRU response calibration. An isolation calibration will
eliminate isolation error (crosstalk error) from the test setup in a transmission test using
that port.
Figure 4-7
2-Port error model (THRU response)
Figure 4-8
2-Port Error model (THRU re isolation)
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
Response (Thru)
.
Step 5.
Press
Select Ports
.
Step 6.
Select the test ports (and corresponding S parameters) upon which a THRU response
calibration is to be performed.
Softkey
Function
2-1 (S21)
Selects test port 2 (input) and test port 1 (output); Corresponds to the
determination of S21
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...