142
Chapter 4
Calibration
3-port TRL calibration
coefficient (by pressing
Done
).
Step 7.
Thru/Line
to start the measurement of the calibration standard.
Softkey
Function
Port x-y Thru
Executes thru/line calibration for test ports x and y.
Port x-z Thru
Executes thru/line calibration for test ports x and z.
Port y-z Thru
Executes thru/line calibration for test ports y and z.
Step 8.
Reflect
to start the measurement of the calibration standard.
Softkey
Function
Port x Reflect
Executes reflection calibration for test port x.
Port y Reflect
Executes reflection calibration for test port y.
Port z Reflect
Executes reflection calibration for test port z.
Step 9.
Line/Match
to start the measurement of the calibration standard.
Softkey
Function
x-y Line/Match
*1.This is equivalent to measuring
x-y Fwd (Syx)
and
x-y Rvs (Sxy)
.
*1
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
x-y Fwd (Syx)
*2.Measuring
x-y Fwd (Syx)
and
x-y Rvs (Sxy)
is equivalent to measuring
x-y Line/Match
.
*2
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement only.
x-y Rvs (Sxy)
Executes line/match calibration for test ports x and y. Use this to perform
reverse-direction measurement only.
x-z Line/Match
Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
x-z Fwd (Szx)
Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement only.
x-z Rvs (Sxz)
Executes line/match calibration for test ports x and z. Use this to perform
reverse-direction measurement only.
y-z Line/Match
Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
y-z Fwd (Szy)
Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement only.
y-z Rvs (Syz)
Executes line/match calibration for test ports y and z. Use this to perform
reverse-direction measurement only.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...