166
Chapter 4
Calibration
Adapter Characterization
6. Measuring Data
Step 1.
Connect the probe to the end of the test fixture on the DUT side.
Step 2.
Depending on the setting of Cal Kit 1, connect each standard to the connector side of the
test fixture and then press the corresponding button (4 in Figure 4-41). When the data
measurement is complete, the button turns yellow.
7. Saving to file
Press the
Save
button (6 in Figure 4-41) to save the calculated S-parameters to a 2-port
Touchstone file.
8. Closing VBA Macro
Press
Close
(7 in Figure 4-41) to close the macro.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...