![Agilent Technologies E5070B Скачать руководство пользователя страница 273](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065273.webp)
Chapter 6
267
Data Analysis
Searching for Positions that Match Specified Criteria
6. Da
ta
Analysis
Figure 6-10
Target search (when transition type is set to “both positive and negative”)
Executing a Target Search
Step 1.
Following Step 1 to Step 4 in “Reading values on the trace” on page 253, activate the
marker you are using for the target search.
Step 2.
Press
.
Step 3.
Press
Target
.
Step 4.
Press
Target Value
and enter the target value in the entry area that appears.
This causes the target search to be executed based on the definitions of the newly set target
value and the currently set transition type.
Step 5.
Press
Target Transition
.
Step 6.
Select a transition type.
Softkey
Function
Positive
Selects Positive as transition type
Negative
Selects Negative as transition type
Both
Selects Both Positive and Negative as transition type
This sets the target search to be executed based on the definitions of the currently set target
value and the newly set transition type.
Step 7.
Press the corresponding softkey to move the marker to the target.
Softkey
Function
Search Target
Executes target search
Search Left
Executes left search
Search Right
Executes right search
NOTE
When the data format is in Smith chart or polar format, execute the search for the main
response value of the two marker response values.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...