![Agilent Technologies E5070B Скачать руководство пользователя страница 660](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065660.webp)
654
Appendix D
Softkey Functions
Analysis Menu
(Continued)
Fixture Simulator
Cmn
ZConversion
Displays softkeys for setting up the common impedance conversion function. When the reference impedance of
two unbalanced ports is expressed as Z
0
, an automatic conversion sets the differential mode reference impedance
of the converted balanced port to 2Z
0
and the common mode reference impedance to Z
0
/2. Common impedance
conversion is a function that further converts Z
0
/2 into another reference impedance.
Cmn
ZConversion
Enables or disables the common impedance conversion function.
:CALC{1-16}:FSIM:BAL:CZC:STAT
Port1 (bal)
Real
Sets the common reference impedance (real format) of balanced port 1.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:Z0
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:REAL
Port1 (bal)
Imag
Sets the common reference impedance (complex format) of balanced port 1.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:IMAG
Port2 (bal) Real
Sets the common reference impedance (real format) of balanced port 2.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR{1-2}:Z0
:CALC{1-16}:FSIM:BAL:CZC:
BPOR{1-2}:REAL
Port2 (bal) Imag
Sets the common reference impedance (complex format) of balanced port 2.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR{1-2}:IMAG
Port3 (bal)
Real
Sets the common reference impedance (real format) of balanced port 3.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:Z0
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:REAL
Port3 (bal)
Imag
Sets the common reference impedance (complex format) of balanced port 3.
:CALC{1-16}:FSIM:BAL:CZC:
BPOR1:IMAG
Return
Returns to the softkey display screen one level higher.
De-Embedding S4P
Displays the softkey to make the setting of the 4-port network embedding/de-embedding feature. The 4-port
network embedding/de-embedding feature is the feature to embed or de-embed your desired network that you have
defined in a 4-port touchstone data file.
De-Embedding
S4P
Toggles ON/OFF the 4-port network embedding/de-embedding feature.
:CALC{1-16}:FSIM:EMB:STAT
Topology
Displays the softkey to assign a topology (the connection method between the analyzer and the DUT).
Select
Topology
Displays the softkey to select the type of the topology (see Figure 7-6).
A
Selects topology A (connecting the 2 ports of the analyzer and the DUT).
:CALC{1-16}:FSIM:EMB:TYPE A
Selects topology B (connecting the 2 ports of the analyzer and the DUT).
:CALC{1-16}:FSIM:EMB:TYPE B
Selects topology C (connecting the 2 ports of the analyzer and the DUT).
:CALC{1-16}:FSIM:EMB:TYPE C
Cancel
Returns to the softkey display screen one level higher.
Ports
Displays the softkey to select a port of the analyzer (test port) that you want to connect to the DUT. (When you
select
A
on the
Select Topology
menu.)
1-2
Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test
ports 1 and 2, respectively.
:CALC{1-16}:FSIM:EMB:TOP:A:PORT
1,2
1-3
Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test
ports 1 and 3, respectively.
:CALC{1-16}:FSIM:EMB:TOP:A:PORT
1,3
1-4
Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test
ports 1 and 4, respectively.
:CALC{1-16}:FSIM:EMB:TOP:A:PORT
1,4
2-1
Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test
ports 2 and 1, respectively.
:CALC{1-16}:FSIM:EMB:TOP:A:PORT
2,1
2-3
Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test
ports 2 and 3, respectively.
:CALC{1-16}:FSIM:EMB:TOP:A:PORT
2,3
Key Operation
Function
SCPI Command
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...