Chapter 4
Calibration
Changing the Calibration Kit Definition
NOTE
Select
Set All
to use the same standards for all test ports.
Step 18.
Select the standards to be registered in the class from among standards numbered 1 through
21. To change the connectors between male and female (e.g. OPEN (f) to OPEN (m)),
select the appropriately labeled standards here.
Step 19.
Repeat Step 17 and Step 18 until classes are defined for all test ports that need to be
redefined.
Step 20.
Press
Return
.
Step 21.
Repeat Step 15 to Step 20 to redefine all classes that need to be modified.
Step 22.
Press
Return
.
Step 23.
Press
Label Kit
and input a new label for the calibration kit by using the keypad displayed
on the screen.
Example of defining the TRL calibration kit
You need to enter the definition of the TRL calibration kit to perform TRL
calibration.Follow these steps to define the following calibration kit given as an example.
•
THRU (Delay 0 ps, Offset Loss 1.3 Gohm/s)
•
REFLECT(SHORT, Delay 0 ps)
•
MATCH (@0 to 2GHz)
•
LINE1 (50-ohm transmission line, Delay 54.0 ps @ 2G to 7GHz)
•
LINE2 (50-ohm transmission line, Delay 13.0 ps @ 7G to 32GHz)
Procedure to define the name of the calibration kit
Step 1.
Press
.
Step 2.
Press
Cal Kit
.
Step 3.
Select a User kit that has not been registered.
Step 4.
Press
Modify Kit
.
Step 5.
Press
Label Kit [User]
.
Step 6.
Type in a name you want.
Procedure to define Thru and Reflect
Step 1.
Press
Define STDs
.
Step 2.
Press
1:No Name - Label
.
Step 3.
Type in
THRU
.
Step 4.
Select
STD Type
-
Delay/Thru
.
Step 5.
Set
Offset Loss
to 1.3Gohm/s, and
Offset Delay
to 0.
Step 6.
Press
Return
to return to the Define STD menu.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...