![Agilent Technologies E5070B Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065137.webp)
Chapter 4
131
Calibration
ECal (electronic calibration)
4. Cal
ibra
tion
Turning off ECal auto-detect function
The ECal module automatically detects the connection between E5070B/E5071B's test
ports and ECal module's ports. You can turn off this function to set ports manually.
NOTE
Even if the connection is wrong with the auto-detect function turned off, no error is
displayed.
Step 1.
Press
or
to select the channel for which you want to turn off the
auto-detect function.
Step 2.
Press
.
Step 3.
Press
ECal
.
Step 4.
Press
Orientation
, and select
Manual
.
Softkey
Function
Auto
Turns on the auto-detect function.
Manual
Turns off the auto-detect function.
Step 5.
Specify a test port of the E5070B/E5071B.
Softkey
Function
Port 1
Specifies port 1 of the E5070B/E5071B.
Port 2
Specifies port 2 of the E5070B/E5071B.
Port 3
*1.Only with Options 313, 314, 413, and 414.
*1
Specifies port 3 of the E5070B/E5071B.
Port 4
*2.Only with Options 413 and 414.
*2
Specifies port 4 of the E5070B/E5071B.
Step 6.
Specify an Ecal port for the port of the E5070B/E5071B you specified.
Softkey
Function
Port A
Specifies port A of Ecal.
Port B
Specifies port B of Ecal.
Port C
*1.Only with Ecal modules supporting 3 ports and 4 ports.
*1
Specifies port C of Ecal.
Port D
*2.Only with Ecal modules supporting 4 ports.
*2
Specifies port D of Ecal.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...