Chapter 15
Measurement Examples
Measuring Parameters with Cable
5. Setting the Time Domain Function
Set the conversion function to display the response in time the domain. If you enable this
setting, the response in time domain is displayed as shown in Figure 15-34. A peak
indicating a small mismatch appears at the location of the connector.
Setting Description
Key Operation
Data format: real
-
Real
Setting the transformation type to low-pass
impulse
-
Transform - Lowpass Imp.
Setting the window type to maximum.
Window
-
Maximum
Setting the display range: from 0 s to 10 ns
Start
-
Stop
-
Enabling the transformation function
Transform
(set to
ON
)
Executing auto scale
-
Auto Scale
Figure 15-34
Response in time domain
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...