Appendix D
723
Softkey Functions
System Menu
D. Sof
tke
y
Fu
nctions
(Continued)
Misc Setup
Control Panel...
Opens a control panel window.
None
Return
Returns to the softkey display screen one level higher.
Backlight
Turns the backlight for the LCD screen on/off.
:SYST:BACK
Firmware Revision
Displays the firmware revision information in a dialog box.
*IDN?
Service Menu
Displays softkeys for maintenance services.
Test Menu
Displays softkeys for self diagnosis functions.
Power On Test
Executes an internal test.
None
Display
Executes a display test.
None
Front Panel
Tests the front panel keys (hardkeys).
None
Adjust Touch
Screen
Executes calibration of the touch screen.
None
Return
Returns to the softkey display screen one level higher.
System Correction
Turns ON/OFF error correction, which uses system calibration data. If
user calibration using [Cal] is executed properly and the error
correction is valid, you can turn off system error correction and reduce
measurement time.
:SYST:CORR
Avoid Spurious
Turns ON/OFF the avoid spurious mode.
:SENS{1-16}:SWE:ASP
High Temperature
Turns ON/OFF the high temperature measurement mode. When the
high temperature measurement mode is turned on, drift error can be
reduced in the ambient temperature range of 28×C to 33×C. If you use
the analyzer within the range of 23×C ± 5×C, you must turn off this
function.
:SYST:TEMP:HIGH
Init Src Ctrl
Turns ON/OFF the initial source port control function.
:SYST:ISPC
Init Src Port
Display softkeys for the initial source port control function.
1
Selects set port to 1.
:SYST:ISPC:PORT 1
2
Selects set port to 2.
:SYST:ISPC:PORT 2
3
Selects set port to 3.
:SYST:ISPC:PORT 3
4
Selects set port to 4.
:SYST:ISPC:PORT 4
Cancel
Returns to the softkey display screen one level higher.
Security Level
Displays softkeys for the security level of the frequency blank display function.
None
Turns off the frequency blank display function.
:SYST:SEC:LEV NON
Low
Turns on the frequency blank display function (security level : LOW).
:SYST:SEC:LEV LOW
High
Turns on the frequency blank display function (security level : HIGH).
:SYST:SEC:LEV HIGH
Cancel
Returns to the softkey display screen one level higher.
Key Operation
Function
SCPI Command
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...