![Agilent Technologies E5070B Скачать руководство пользователя страница 398](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065398.webp)
392
Chapter 11
Limit Test
Bandwidth Test
•
Status register (for further information, see the
Programmer's Guide
)
Set up bandwidth test
You must set up the bandwidth threshold and the upper and lower limits before you can use
the bandwidth test function. You can specify the threshold, upper limit, and lower limit for
each trace.
Operational procedure
Follow the steps below to set up the bandwidth test.
Step 1.
Press
or
to activate the channel to which you want to apply the
bandwidth test function.
Step 2.
Press
or
to activate the trace to which you want to apply the
bandwidth test function.
Step 3.
Press
to display the Analysis menu.
Step 4.
Press
Bandwidth Limit
to display the softkeys for the bandwidth test.
Step 5.
Press
N dB Points
to specify the bandwidth threshold. The unit is dB.
Step 6.
Press
Min Bandwidth
to enter the lower limit for the bandwidth. Similarly, press
Max
Bandwidth
to enter the upper limit for the bandwidth. The unit is Hz for both Min and Max
bandwidths.
NOTE
If the data format is Smith chart or polar, the limit test is performed for the main response
value among the two marker response values.
Turning On/Off Bandwidth Test and Displaying Results
You can configure the on/off setting of the bandwidth test function for each trace
independently.
Operational procedure
Follow the steps below to configure the on/off setting of the bandwidth test.
Step 1.
Press
or
to activate the channel to which you want to apply the
bandwidth test function.
Step 2.
Press
or
to activate the trace to which you want to apply the
bandwidth test function.
Step 3.
Press
to display the Analysis menu.
Step 4.
Press
Bandwidth Limit
to display the softkeys for the bandwidth test.
Step 5.
Press
BW Test
and turn on the bandwidth test. You can display the bandwidth marker on
the screen by turning on
BW Marker
. To display the bandwidth value, turn on
BW Display
.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...