Appendix D
673
Softkey Functions
Calibration Menu
D. Sof
tke
y
Fu
nctions
(Continued)
Calibrate
Adapter Removal
Displays softkeys for executing adapter removal/insertion.
Select Port
Displays softkeys for selecting a test port.
1
Selects test port 1.
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
1
2
Selects test port 2.
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
2
3
Selects test port 3.
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
3
4
Selects test port 4.
:SENS{1-16}:CORR:COLL:METH:ADAP:REM
4
Cancel
Returns to the softkey display screen one level higher.
Cal Kit
Displays softkeys for selecting calibration kits.
:SENS{1-16}:CORR:COLL:CKIT
Open
Executes an OPEN calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:OPEN
Short
Executes a SHORT calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:SHOR
Load
Executes a LOAD calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:LOAD
Adapter Length
Sets the adapter length on the test port selected in the
Select Port
menu.
:SENS{1-16}:CORR:COLL:ADAP{1-4}:LENG
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
Key Operation
Function
SCPI Command
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...