![Agilent Technologies E5070B Скачать руководство пользователя страница 332](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065332.webp)
326
Chapter 8
Frequency-Offset Measurement (Option 008)
Measurement of Harmonic Distortion
Measurement of Harmonic Distortion
Using the frequency-offset function and absolute measurement function in combination
allows you to measure harmonic distortion of nonlinear devices such as mixers and
amplifiers.
Measurement flow
Table 8-4
Measurement Flow of Harmonic Distortion
Item
Description
"1. Setting
Frequency-Offset" on page
313
Sets frequency-offset function
"2. Implementing Receiver
Calibration" on page 327
Implements receiver calibration
"3. Setting Absolute
Measurement Parameters"
on page 327
Sets absolute measurement parameters
•
Procedures for setting absolute measurement parameters
"4. Harmonic Distortion
Measurement" on page 328
Implements harmonic distortion measurement
•
Connection of DUT
•
Procedures for setting measurement parameters
Table 8-4 shows the measurement flow.
1. Setting Frequency-Offset Function
The frequency-offset function allows you to make measurements while the frequencies are
different at each test port. In this case, measurement is done by using the setting example
for measurement frequency in Figure 8-12. For the setting of frequency-offset sweep, see
"1. Setting Frequency-Offset" on page 313.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...