610
Appendix B
Troubleshooting
Troubleshooting
•
The touch screen becomes inoperable.
³
Using the front panel keys, turn
System
–
Key Lock
–
Touch Screen & Mouse
Lock
OFF.
³
Execute the calibration of the touch screen. For information on the execution
procedure, see “Calibration of the Touch Screen” on page 470.
•
The mouse becomes inoperable.
³
Using the front panel keys, turn
System
–
Key Lock
–
Touch Screen & Mouse
Lock
OFF.
•
All of the front panel keys, keyboard, and mouse become inoperable.
³
Confirm that the keyboard or mouse is connected correctly. When it is connected
correctly, turn off the power once, and restart the system.
•
The keyboard and mouse have been connected after power-on.
³
Turn off the power once, and restart the system.
When taking all these measures does not recover operability, there is a possibility of a
failure. Contact Agilent Technology’s Customer Contact listed at the end of this guide or
the company from which you bought the device.
The Screen Freezes and All Operations Become Impossible.
•
The measurement in progress or screen update is stalled and all of the front panel keys,
keyboard, mouse, and touch screen (Option 016) are inoperable.
³
Press the standby switch to turn off the power once, and restart the system.
When a similar symptom reappears, there is a possibility of a failure. Contact Agilent
Technology’s Customer Contact listed at the end of this guide or the company from which
you bought the device.
The System Freezes while in Operation.
•
The system freezes while in operation.
³
Press the standby switch to turn off the power once, and restart the system.
The Rear Cooling Fan Does Not Operate.
There is a possibility of a failure. Contact Agilent Technology’s Customer Contact listed at
the end of this guide or the company from which you bought the device.
You Cannot Save a File in a Floppy Disk.
•
You cannot save a file in a floppy disk.
³
Confirm that the floppy disk is initialized. If not, initialize it.
³
Confirm that the floppy disk is inserted correctly. Insert a floppy disk until the eject
button pops up fully.
³
Confirm that the floppy disk is not write-protected. If it is, unprotect the disk.
³
Confirm that the floppy disk has free space. If it does not, delete unnecessary files or
use a new floppy disk.
When taking all these measures does not make it possible to save a file, there is a
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...