Chapter 6
263
Data Analysis
Searching for Positions that Match Specified Criteria
6. Da
ta
Analysis
Searching for Positions that Match Specified Criteria
You can search for a position that matches your specified criteria by using the Marker
Search feature.
Marker Search allows you to search for a position that matches any of the following
criteria.
R
Maximum value
R
Minimum value
R
Target (a point that has a target measurement value)
•
Target nearest to marker position
•
Target on left-hand side nearest to marker position
•
Target on right-hand side nearest to marker position
R
Peak
•
Maximum peak (for a positive peak), minimum peak (for a negative peak)
•
Peak on left-hand side nearest to marker position
•
Peak on right-hand side nearest to marker position
Setting search range
The Marker Search feature allows you to set part of the sweep range as the search target
(Partial Search feature) as well as the entire search range. For the Partial Search feature,
you can select whether to couple traces in the channel.
Procedure to Turn On/Off Trace Coupling within Search Range
Step 1.
Press
or
to activate the channel for which you want to set the
search range.
Step 2.
Press
or
to activate the trace for which you want to set the search
range.
Step 3.
Press
.
Step 4.
Press
Search Range
.
Step 5.
Press
Couple
to toggle on/off trace coupling within the search range.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...