Figure 4-55 and Figure 4-56 show the parameters used in defining standards.
Figure 4-55
Reflection Standard Model (SHORT, OPEN, or LOAD)
Figure 4-56
Transmission Standard Model (THRU)
Z0
The offset impedance between the standard to be defined and the
actual measurement plane. Normally, this is set to the system’s
characteristic impedance.
Delay
The delay that occurs depending on the length of the transmission line
between the standard to be defined and the actual measurement plane.
In an OPEN, SHORT, or LOAD standard, the delay is defined as
one-way propagation time (sec.) from the measurement plane to the
standard. In a THRU standard, it is defined as one-way propagation
time (sec.) from one measurement plane to the other. The delay can be
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...