Chapter 4
109
Calibration
OPEN/SHORT Response Calibration (reflection test)
4. Cal
ibra
tion
Step 8.
Press
Open
or
Short
to start the calibration measurement.
Step 9.
If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a.
Connect a LOAD standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Figure 4-6
Connecting the LOAD standard
b.
Press
Load (Optional)
to start the measurement on the LOAD standard.
Step 10.
Press
Done
to terminate the response calibration (and the LOAD isolation calibration)
process. Upon pressing this key, calibration coefficients will be calculated and saved. The
error correction function will also be automatically enabled.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...