Chapter 4
111
Calibration
THRU Response Calibration (transmission test)
4. Cal
ibra
tion
NOTE
An asterisk (*) in the upper right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
calibration coefficient (pressing
Done
).
Step 7.
Make a THRU connection between the test ports (between the connectors to which the
DUT will be connected) selected in Step 6.
Figure 4-9
Connecting at THRU response calibration
3-1 (S31)
Selects test port 3 (input) and test port 1 (output); Corresponds to the
determination of S31
4-1 (S41)
Selects test port 4 (input) and test port 1 (output); Corresponds to the
determination of S41
1-2 (S12)
Selects test port 1 (input) and test port 2 (output); Corresponds to the
determination of S12
3-2 (S32)
Selects test port 3 (input) and test port 2 (output); Corresponds to the
determination of S32
4-2 (S42)
Selects test port 4 (input) and test port 2 (output); Corresponds to the
determination of S42
1-3 (S13)
Selects test port 1 (input) and test port 3 (output); Corresponds to the
determination of S13
2-3 (S23)
Selects test port 2 (input) and test port 3 (output); Corresponds to the
determination of S23
4-3 (S43)
Selects test port 4 (input) and test port 3 (output); Corresponds to the
determination of S43
1-4 (S14)
Selects test port 1 (input) and test port 4 (output); Corresponds to the
determination of S14
2-4 (S24)
Selects test port 2 (input) and test port 4 (output); Corresponds to the
determination of S24
3-4 (S34)
Selects test port 3 (input) and test port 4 (output); Corresponds to the
determination of S34
Softkey
Function
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...