Chapter 4
Calibration
Adapter Removal-Insertion
Procedure for Adapter Removal / Insertion
The s-parameter of a reciprocal adapter can be determined when the following data is
available.
•
Open, Short, and Load measurements.
•
Actual values derived from the CalKit definitions.
•
An approximate length of the adapter.
•
Nature of the intended operation : removal or insertion.
Once the s-parameter has been determined, run de-embedding and update the error
coefficients.
NOTE
If the frequency setting is not zero span, an approximate length of the adapter should be
provided otherwise an error generated in the measurement.
Step 1.
Perform a full n-port calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
Adapter Removal
.
Step 5.
Select the port in/from which you want to insert/remove adapter characteristics from
Select Port
.
Step 6.
Press
Cal Kit
, and then select calibration kit.
Step 7.
Select the adapter length.
Softkey
Function
AUTO
Adapter automatic
Positive Value
Adapter removal
Negative Value
Adapter insertion
NOTE
For adapter removal, the adapter length is positive . For adapter insertion, the adapter
length is negative.
Step 8.
Press
Open
,
Short
, and
Load
respectively.
Step 9.
Press
Done
to complete the process.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...