![Agilent Technologies E5070B Скачать руководство пользователя страница 544](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065544.webp)
Table 16-4
Corrected System Performance With 3.5 mm Device Connector Type, 85033E
Calibration Kit
Network analyzer: E5070B/E5071B, Calibration kit: 85033E (3.5 mm, 50
Ω
), Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23
°
C
±
5
°
C with <1
°
C deviation
from calibration temperature, Isolation calibration not omitted
Description
Specification (dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity
46
38
38
Source Match
43
37
36
Load Match
46
38
38
Reflection Tracking
±
0.006
±
0.009
±
0.010
Transmission Tracking
±
0.016
±
0.065
±
0.079
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
538
Chapter 16
Specifications and Supplemental Information
Corrected System Performance
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...