Appendix D
665
Softkey Functions
Calibration Menu
D. Sof
tke
y
Fu
nctions
(Continued)
Calibrate
Enhanced
Response
Select Port
3-1 (S31
S11)
Selects the transmission test (measurement of S
31
) for test port
1
→
3.
:SENS{1-16}:CORR:COLL:METH:ERES 3,1
4-1 (S41
S11)
Selects the transmission test (measurement of S
41
) for test port
1
→
4.
:SENS{1-16}:CORR:COLL:METH:ERES 4,1
1-2 (S12 S22)
Selects the transmission test (measurement of S
12
) for test port
2
→
1.
:SENS{1-16}:CORR:COLL:METH:ERES 1,2
3-2 (S32
S22)
Selects the transmission test (measurement of S
32
) for test port
2
→
3.
:SENS{1-16}:CORR:COLL:METH:ERES 3,2
4-2 (S42
S22)
Selects the transmission test (measurement of S
42
) for test port
2
→
4.
:SENS{1-16}:CORR:COLL:METH:ERES 4,2
1-3 (S13
S33)
Selects the transmission test (measurement of S
13
) for test port
3
→
1.
:SENS{1-16}:CORR:COLL:METH:ERES 1,3
2-3 (S23
S33)
Selects the transmission test (measurement of S
23
) for test port
3
→
2.
:SENS{1-16}:CORR:COLL:METH:ERES 2,3
4-3 (S43
S33)
Selects the transmission test (measurement of S
43
) for test port
3
→
4.
:SENS{1-16}:CORR:COLL:METH:ERES 4,3
1-4 (S14
S44)
Selects the transmission test (measurement of S
14
) for test port
4
→
1.
:SENS{1-16}:CORR:COLL:METH:ERES 1,4
2-4 (S24
S44)
Selects the transmission test (measurement of S
24
) for test port
4
→
2.
:SENS{1-16}:CORR:COLL:METH:ERES 2,4
3-4 (S34
S44)
Selects the transmission test (measurement of S
34
) for test port
4
→
3.
:SENS{1-16}:CORR:COLL:METH:ERES 3,4
Cancel
Returns to the softkey display screen one level higher.
Open
Executes an OPEN calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:OPEN
Short
Executes a SHORT calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:SHOR
Load
Executes a LOAD calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:LOAD
Thru
Executes a THRU response calibration on the test ports selected in
the
Select Ports
menu.
:SENS{1-16}:CORR:COLL:THRU
Isolation
(Optional)
Executes an isolation calibration on the test ports selected in the
Select Ports
menu.
:SENS{1-16}:CORR:COLL:ISOL
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
Key Operation
Function
SCPI Command
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...