Appendix F
745
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
F. Re
pla
cin
g th
e 87
53
ES
with th
e E507
0B/E507
1B
Comparing Functions
Table F-5
Functions of the 8753ES vs. the E5070B/E5071B
Function
8753ES
E5070B/E5071B
Measurement Reset
Can be executed by using the front
panel and the GPIB command.
Can be executed by using the front
panel, the GPIB command, and
telnet.
Channel
Number of channels
4 channels
4/9/12/16 channels
Coupling between channels
Channels 3 and 4 are auxiliary
channels for channels 1 and 2 and
subject to the same sweep
conditions, etc. applicable to
channels 1 and 2.
Couple/Uncouple between channels
1 and 2 can be set up.
Channels are independent of each
other.
Trace
A trace for each channel
Each channel can accommodate up
to 4/9/12/16 traces.
Measurement parameter
S-parameters, A, B, R, A/R, B/R,
A/B, and analog bus
S-parameters (Mixed-mode
S-parameters),
Absolute value meas parameter
S-parameter conversion
Impedance (reflection and
transmission), admittance (reflection
and transmission), and 1/S
←
Display format (Data format)
log magnitude, linear magnitude,
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
log magnitude, linear magnitude,
phase, expanded phase, positive
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
Test port output Number of ports
2 ports
2 ports (Opts. 213/214)
3 ports (Opts. 313/314)
4 ports (Opts. 413/414)
Frequency
30 kHz to 3 GHz (Std.)
30 kHz to 6 GHz (Opt. 006)
300 kHz to 3 GHz (E5070B)
300 kHz to 8.5 GHz (E5071B)
Power level
-85 to +10 dBm (Std.)
-85 to +8 dBm (Opts. 014/075)
-15 to 10 dBm (Opts. 213, 313, 413)
-50 to 10 dBm (Opts. 214, 314, 414)
Characteristic Impedance
50
Ω
(Std.)
75
Ω
(Opt. 075)
50
Ω
Coupled/Uncoupled power
levels
You can define Couple/Uncouple
between channels and between ports.
Traces are coupled on the same
channel, not between channels.
Power slope function
Available
←
Turning the output On/Off
Allowed
←
Power range
Items to be set
up
Per port and per channel
Per channel (Opts. 214/314/414)
If the above options are not installed,
the default range is -15 to 10 dBm.
Automatic
setting
On/Off setting capability
Manual setting
Setting range
Range setting with 10 dBm
resolution is definable from between
-15 and +10 dBm down to between
-85 and -60 dBm.
Range setting with 5 dBm resolution
is definable from between -15 and 10
dBm down to between -50 and -25
dBm
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...