746
Appendix F
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
Measurement
(cont'd.)
Sweep
conditions
Number of points
(except for the list frequency
sweep operation)
3, 11, 21, 26, 51, 101, 201, 401, 801,
1601
Arbitrary value from 2 to 1601
*1
Sweep type
linear sweep, log sweep, list sweep,
power sweep, and CW time sweep
linear sweep, log sweep, segment
sweep, and power sweep
List frequency
sweep
Number of list
tables
One for channels 1 and 3 and one for
channels 2 and 4, two in total
One for each channel (16 in total)
Number of
segments per
table
Up to 30
Up to 201
Creation of a
table using
GPIB(SCPI)
command
Creating a table using more than one
command
Creating a table using a single
command
Number of
points
1 to 1601 points per segment
Up to 1601 points in total
2 to 1601 points per segment
Stepped/swept
mode
stepped mode and swept mod.
Selectable only when list frequency
sweep is performed
stepped mode, swept mode, fast
stepped mode, and fast swept mode.
Selectable also when list frequency
sweep is not performed.
Sweep of
designated
segments.
Allowed
Not allowed (Always sweeps all
segments.)
Sweep time
Automatic, Manual (definable from
the shortest time to 24 hours)
Automatic, Manual (range of
definable sweep depends on sweep
condition)
IF bandwidth
10 Hz, 30 Hz, 100 Hz, 300 Hz,
1 KHz, 3 kHz, 3.7 kHz, 6 kHz
10 Hz, 15 Hz, 20 Hz, 30 Hz,
40 Hz, 50 Hz, 70 Hz, 100 Hz,
150 Hz, 200 Hz, 300 Hz,
400 Hz, 500 Hz, 700 Hz, 1 kHz,
1.5 kHz, 2 kHz, 3 kHz, 4 kHz,
5
K
H
Z
, 7
K
H
Z
, 10
K
H
Z
, 15
K
H
Z
,
20
K
H
Z
, 30
K
H
Z
, 40
K
H
Z
, 50
K
H
Z
,
70
K
H
Z
, 100 kHz
Averaging
Can be set from 1 to 999.
Can be set from 1 to 999.
Smoothing
The smoothing aperture can be set
from 0.05% to 20%.
The smoothing aperture can be set
from 0.05% to 25%.
Electrical delay Transmission line
Compatible with coaxial cables and
waveguides
Compatible with coaxial cables only
Value
Any value from 0 to 10 seconds
Any value from -10 to +10 seconds
Phase offset
Any value from -360
°
to +360
°
.
Any value from -360
°
to 360
°
.
Setting sweep conditions for Couple/Uncouple
Channels 1 and 3 and channels 2 an 4
are always coupled. Channels 1 and
2 can be set either at Couple or
Uncouple.
Traces in the same channel are
coupled; traces in different channels
are not coupled.
Trigger mode
Continuous, single, hold, specified
number of sweeps
Continuous, single, hold
Table F-5
Functions of the 8753ES vs. the E5070B/E5071B
Function
8753ES
E5070B/E5071B
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...