![Agilent Technologies E5070B Скачать руководство пользователя страница 411](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065411.webp)
Chapter 12
405
Optimizing Measurements
Setting Port Extensions and Loss Values
1
2
. O
pt
imiz
ing
M
e
as
ur
em
e
nt
s
Step 5.
Method
to set the frequencies used for calculation.
Softkey
Function
Current Span
Executed using the frequency range set currently.
Active Marker
Executed using the frequency at the active marker.
*1.Even if the active marker has been disabled, it is automatically enabled.
*1
In this case, the result
is applied to Loss1. Loss2 is ignored.
User Span
Executed using a start value and a stop value you set.
Step 6.
If you have selected
User Span
, use
User Span Start
and
User Span Stop
to set a start
value and a stop value.
NOTE
For
Current Span
and
User Span
, a frequency point at 1/4 of the frequency range is set to
Freq1
; a frequency point at 3/4 of the frequency range is set to
Freq2
NOTE
If the setting is not made before starting OPEN/SHORT standard measurement, it does not
affect the calculation result.
Figure 12-5
Frequency points in a frequency range (Freq1 and Freq2)
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...