602
Appendix A
Manual Changes
Manual Changes
o Assignable x-axis such as RF+LO, RF-LO, and LO-RF frequencies for each active
trace
o Conjugation for converting vector mixer measurement parameters
o 7 mm calibration kits such as 85031B and 85050C/D
o Calibration data and calibration coefficient clear functions
Change 6
Use Following specifications for the serial prefix MY422.
Table A-3
Description
Specification
Supplemental Information
Damage Level
300 kHz to 8.5 GHz
RF +20 dBm,
±
14 VDC (Source attenuator = 0
dB),
±
25 VDC (Source attenuator = 5
dB or more), typical
Change 5
Use Following specifications for the serial prefix MY421 and JP1KK.
Table A-4
System Dynamic Range
Description
Specification
Supplemental Information
System Dynamic Range
*1*2
300 kHz to 3 MHz
IF bandwidth = 3 kHz
85 dB
3 MHz to 1.5 GHz
95 dB
98 dB
1.5 GHz to 4 GHz
97 dB
100 dB
4 GHz to 6 GHz
93 dB
95 dB
6 GHz to 7.5 GHz
88 dB
91 dB
7.5 GHz to 8.5 GHz
81 dB
84 dB
300 kHz to 3 MHz
IF bandwidth = 10 Hz
110 dB
3 MHz to 1.5 GHz
120 dB
123 dB
1.5 GHz to 4 GHz
122 dB
125 dB
4 GHz to 6 GHz
118 dB
120 dB
6 GHz to 7.5 GHz
113 dB
116 dB
7.5 GHz to 8.5 GHz
106 dB
109 dB
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...