514
Chapter 15
Measurement Examples
Measuring an Unbalanced and Balanced Bandpass Filter
Step 4.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 3.
Setting Description
Key Operation
Unbalanced-balanced conversion of trace 3:
ON
-
-
Fixture Simulator
-
BalUn
(turns it
ON
)
Measurement parameter: S
ss11
-
Sss11
Data format: Smith chart (marker display:
R+jX)
-
Smith
-
R + jX
Step 5.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 4.
Setting Description
Key Operation
Unbalanced-balanced conversion of trace 4:
ON
-
-
Fixture Simulator
-
BalUn
(turns it
ON
)
Measurement parameter: S
dd22
-
Sdd22
Data format: Smith chart (marker display:
R+jX)
-
Smith
-
R + jX
Figure 15-28 shows the setting results for each parameter.
Figure 15-28
Measurement results after unbalanced-balanced conversion
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...