Appendix D
667
Softkey Functions
Calibration Menu
D. Sof
tke
y
Fu
nctions
(Continued)
Calibrate
2-Port Cal
Transmission
Displays softkeys for executing transmission calibrations. Symbols x and y refer to the two test ports selected in
the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x-y Thru
Executes a THRU calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:THRU x,y
:SENS{1-16}:CORR:COLL:THRU y,x
Return
Returns to the softkey display screen one level higher.
Isolation
(Optional)
Displays softkeys for executing isolation calibrations. Symbols x and y refer to the two test ports selected in the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x-y Isol
Executes an isolation calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:ISOL x,y
:SENS{1-16}:CORR:COLL:ISOL y,x
Return
Returns to the softkey display screen one level higher.
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Overwrite
Terminates partial overwrite and recalculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:PART:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
3-Port Cal
Displays softkeys for executing full 3-port calibrations.
Select Ports
Displays softkeys for selecting test ports.
1-2-3
Selects test ports 1, 2, and 3.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,2,3
1-2-4
Selects test ports 1, 2, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,2,4
1-3-4
Selects test ports 1, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,3,4
2-3-4
Selects test ports 2, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 2,3,4
Cancel
Returns to the softkey display screen one level higher.
Reflection
Displays softkeys for executing reflection calibrations. Symbols x, y, and z refer to the three test ports selected in
the
Select Ports
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Port x Open
Executes an OPEN reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:OPEN x
Port x Short
Executes a SHORT reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:SHOR x
Port x Load
Executes a LOAD reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:LOAD x
Port y Open
Executes an OPEN reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:OPEN y
Port y Short
Executes a SHORT reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:SHOR y
Port y Load
Executes a LOAD reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:LOAD y
Port z Open
Executes an OPEN reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:OPEN z
Port z Short
Executes a SHORT reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:SHOR z
Port z Load
Executes a LOAD reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:LOAD z
Return
Returns to the softkey display screen one level higher.
Transmission
Displays softkeys for executing transmission calibrations. Symbols x, y, and z refer to the three test ports selected
in the
Select Ports
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Port x-y Thru
Executes a THRU calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:THRU x,y
:SENS{1-16}:CORR:COLL:THRU y,x
Port x-z Thru
Executes a THRU calibration on test ports x and z.
:SENS{1-16}:CORR:COLL:THRU x,z
:SENS{1-16}:CORR:COLL:THRU z,x
Key Operation
Function
SCPI Command
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...